NorthStar Order Entry supports scales that use the "NCI Weigh-Tronics" protocol. It is known to work with the following models:
- Brecknell 6710U
- Mettler Toledo Ariva-S
Table of Contents
General Instructions
Adjust the level
Before installing or connecting cables, adjust the level on the scale.
- Lift the metal plate on the top of the scales to expose the built-in level for the scale.
- Brecknell 6710U - Level
- Mettler Toledo Ariva-S - Level
- Adjust the leg stands on each of the scales until the small bubble in the level is in the center of the circle.
Connect units and cables
- Each scale will have its own display unit and cables provided.
- Follow the manufacturer's manual to properly attach the:
- display unit
- power cable
- serial cable
3. These are the connection ports on the sides of the scales:
Brecknell 670U - Ports | Mettler Toledo Ariva-S - Ports |
4. After the cables and display are connected, connect the scales to the Lantronix device to provide it a COM #. You will enter this COM # in ECM.
You can follow the Lantronix document: Installing a Lantronix Serial-to-Ethernet Device
Troubleshooting: NorthStar Order Entry: Lantronix + Scale data inconsistency troubleshooting
Device-specific information
Brecknell 6710U
The Brecknell 6710U scale uses a straight-through serial cable. It does not work with a null modem cable.
Brecknell 6710U configuration
- User
- COM2
- BudRt (Baud Rate): "9600"
- Bt.Fmt (Bit Format): "7E1"
- OutMd (output mode): "Cmd" or "Prt.Cd"
- Lyout (Layout): SCP01
- COM2
Mettler Toledo Viva
The Mettler Toledo Viva scale uses a null modem cable. It does not work with a straight-through serial cable.
Mettler Toledo Viva Configuration
Setup
This mode allows an operator to change certain soft switches directly from the keyboard. While turning on the
power, press and hold >0< key until “grP 1” is displayed. The scale is configurable for your specific needs
with a special interactive set up procedure. In order to access the various prompts, you must utilize the
following keys during the scale setup mode.
>0<
This key is used to confirm the choice and then go to the next step.
>T<
This key is used to step through the set up groups. Once a group is accepted, this key is
used to select the soft switch settings
Make sure to follow these settings for the scale within the GROUPS in the setup.
GRP1:
1.5: OFF
GRP2:
2.4: 2
- GRP3:
- 3.1: 3
- 3.2: 1
- 3.3: 0
- 3.4: 1
- 3.5: 8
- Save Settings by pressing >0<