Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  • Brecknell 6710U 
  • Mettler Toledo Ariva-S

Table of Contents

Table of Contents
maxLevel2
excludeTable of Contents

General Instructions

 

 


Adjust the level

Before installing or connecting cables, adjust the level on the scale. 

  1. Lift the metal plate on the top of the scales to expose the built-in level for the scale. 
  2. Brecknell 6710U - Level
  3. Mettler Toledo Ariva-S - Level
  4. Adjust the leg stands on each of the scales until the small bubble in the level is in the center of the circle. 


Connect units and cables

  1. Each scale will have its own display unit and cables provided.
  2. Follow the manufacturer's manual to properly attach the: 

...

You can follow the Lantronix document: Installing a Lantronix Serial-to-Ethernet Device

...


Troubleshooting: NorthStar Order Entry: Lantronix + Scale data inconsistency troubleshooting 


Device-specific information

Brecknell 6710U

The Brecknell 6710U scale uses a straight-through serial cable. It does not work with a null modem cable.

Brecknell 67xx Serial Scale User Manual.pdf

Brecknell 6710U configuration

  • User
    • COM2
      • BudRt (Baud Rate): "9600"
      • Bt.Fmt (Bit Format): "7E1"
      • OutMd (output mode): "Cmd" or "Prt.Cd"
      • Lyout (Layout): SCP01

Mettler Toledo Viva

The Mettler Toledo Viva scale uses a null modem cable. It does not work with a straight-through serial cable.

Mettler Toledo Viva Configuration (Viva PDF Document attached to this page)

  • Setup
    This mode allows an operator to change certain soft switches directly from the keyboard. While turning on the
    power, press and hold >0< key until “grP 1” is displayed. The scale is configurable for your specific needs
    with a special interactive set up procedure. In order to access the various prompts, you must utilize the
    following keys during the scale setup mode.
    >0<
    This key is used to confirm the choice and then go to the next step.
    >T<
    This key is used to step through the set up groups. Once a group is accepted, this key is
    used to select the soft switch settings

...


  • Make sure to follow these settings for the scale within the GROUPS in the setup.

    • GRP1

      • 1.5: OFF

    • GRP2:

      • 2.4: 2

    • GRP3:
      • 3.1: 3
      • 3.2: 1
      • 3.3: 0
      • 3.4: 1
      • 3.5: 8
    • Save Settings by pressing >0<

...